6 Maintenance

■■■■■■■■

27

4 Inserting the lamp holder

■

First, push the white lamp holder carefully all the way into the
opening,

■

then turn it approx. 45° clockwise and

■

then slowly let it go.

5 Closing the lamp compartment

■

Place the sealing plate onto the opening. Make sure that no
cables are being pinched.

■

Insert the four screws and tighten them in crosswise sequence
using the hex key.

■

Push the hex key into the guide rail on the cover.

■

Close the lid.

6 Calibrating the instrument

The instrument has to be recalibrated every time a lamp is replaced.

■

Switch on the instrument.

■

Recalibrating the instrument (see step 6 in the chapter "Non-reg-
ulated industries", page 12)
.