6.2 User maintenance

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

6.2.2

Replacing the fan filter

The fan filter has to be checked at least once a month. If the instrument is
operated in a dusty or otherwise dirty environment, then a check is
required once or even twice a week.

The fan is located on the rear of the instrument. The filter cover is
attached to the fan filter by means of four plastic latches.

Required accessories

■

Fan filter (6.7430.020)

1 Switching off the instrument

■

Turn the on/off switch (2-1) to the position O.

2 Removing the filter cover

Grab the filter cover with both hands and gently take it off starting
on top and then proceeding to the sides.

3 Checking the state of the filter

Take out the filter and inspect it carefully.

■

If the filter shows no cracks, then it can be cleaned and used
again.

■

If the filter is damaged, then a new filter has to be used.