6 Troubleshooting

881 Compact IC pro – Anion

■■■■■■■■

93

Problem

Cause

Remedy

Poor reproducibility
of the retention
times

Eluent – Leak in the eluent
path.

Check the eluent path

Eluent – Blockage in the
eluent path.

Check the eluent path

Suppressor – No (or
insufficient) pump-
ing of regeneration
or rinsing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set the
contact pressure", page 49)
.

Peristaltic pump – filter
blocked .

Replace the filter (see "Replacing the filter",
page 81)
.

Suppressor – backpressure
too high.

Clean the suppressor (see Chapter 5.11.3.3,
page 85)
or replace parts
(see Chapter
5.11.3.4, page 86)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing.

Chromatograms
have poor resolution

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.12.4, page 89)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 57)
.

Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)

Capillary connections –
dead volume in the system.

Check connections (see Chapter 3.5, page 17)
(use PEEK capillaries with an internal diameter
of 0.25 mm between the injection valve and
detector).

Guard column – diminished
performance.

Replace guard column (see Chapter 3.18,
page 54)
.

Separation column – dead
volume at column head.

Install the separation column in reverse
flow direction (if permitted by the leaflet)
and rinse into a beaker.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 57)
.

Greater rise in the
baseline

Suppressor – reduced
capacity.

Regenerate the suppressor (see Chapter
5.11.3.2, page 83)
.