Figure 27, Suppressor – connection capillaries – Metrohm 881 Compact IC pro – Anion User Manual
Page 61

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
881 Compact IC pro – Anion
■■■■■■■■
51
1
3
2
3
4
5
6
7
8
2
1
Figure 27
Suppressor – connection capillaries
1
Union nut
2
Connecting piece (6.2832.010)
3
Eluent inlet capillary
Labeled with in.
4
Eluent outlet capillary
Labeled with out.
5
Rinsing solution inlet capillary
Labeled with rinsing solution.
6
Rinsing solution outlet capillary
Labeled with waste rins..
7
Regeneration solution outlet capillary
Labeled with waste reg..
8
Regeneration solution inlet capillary
Labeled with regenerant.
The rinsing solution and regeneration solution are pumped with a peristal-
tic pump (see Chapter 3.15, page 44).
Caution
To protect the suppressor against foreign particles or bacterial growth,
a pump tubing connection with filter (6.2744.180) (25-3) must be
mounted between the peristaltic pump and the inlet capillaries of the
suppressor.