Metrohm 711 Liquino User Manual

Page 118

Advertising
background image

4.8 SMPL mode - fully automatic sampling

711 Liquino, Instructions for use

112

Add auxiliary solution

add

aux. solution


With further Dosinos up to three different auxiliary

solutions can again be added, see '>SMPL Dosing
units'. After the addition of each auxiliary solution a
waiting period can be programmed (in HH:MM:SS).
The addition of auxiliary solutions is optional. It is
possible before or after sample transfer.

Start/wait for determination

start

determination

await

end of

determination

<< Ready

START >>

The Liquino starts the measuring or titration instru-

ment by a start signal via a remote lead or by any
command via a serial RS232 connection. During the
determination the Liquino remains in the ready condi-
tion.

At the end of the determination the measuring or

titration instrument sets a 'Ready' signal via a remote
lead or sends a 'Ready' message via the RS232 inter-
face. When this ready message is received the SMPL
mode is terminated.

Advertising