4.8 SMPL mode - fully automatic sampling
711 Liquino, Instructions for use
112
Add auxiliary solution
add
aux. solution
•
With further Dosinos up to three different auxiliary
solutions can again be added, see '>SMPL Dosing
units'. After the addition of each auxiliary solution a
waiting period can be programmed (in HH:MM:SS).
The addition of auxiliary solutions is optional. It is
possible before or after sample transfer.
Start/wait for determination
start
determination
await
end of
determination
<< Ready
START >>
•
The Liquino starts the measuring or titration instru-
ment by a start signal via a remote lead or by any
command via a serial RS232 connection. During the
determination the Liquino remains in the ready condi-
tion.
•
At the end of the determination the measuring or
titration instrument sets a 'Ready' signal via a remote
lead or sends a 'Ready' message via the RS232 inter-
face. When this ready message is received the SMPL