Metrohm 711 Liquino User Manual

Page 33

Advertising
background image

3 Introduction

711 Liquino, Instructions for use

27

If you want, you can configure your Liquino so that after each dispensing

process the dosing cylinder is refilled automatically.

CONFIG

A

2x

J













<ENTER>

Press the <CONFIG> key and the <> key twice.

Di spl ay:

conf i gur at i on

>manual

Here you will find the submenu with the settings which are effective for

manual functions.

Di spl ay:

>manual DOS1

← dosing connection

dos. uni t : def aul t

In the first line, above right, the address of the dosing connection to

which the particular setting applies is shown. You can make different
settings for each of the four dosing connections. The settings for the
other dosing connections DOS2 to DOS4 are also found in this sub-
menu. Now press the <ENTER> key twice.

Di spl ay:

>manual DOS1

aut o f i l l : of f

The colon after the word 'fill' shows that a list of entries is available for

this menu item. In this case there are only two possibilities: 'on' and
'off'.

DOS NR

SELECT

P

<ENTER>

Press the <SELECT> key in order to switch on automatic filling of the

dosing unit. Confirm the entry with <ENTER>.

2x

QUIT

NO

Exit the configuration menu by pressing <QUIT> twice.

START

Z

Now dispense again by pressing the <START> key and you will see that

the dosing cylinder is automatically refilled after each dosing process.

MANUAL

MAN OFF

D

Pressing the <MAN. OFF / MANUAL> key switches manual operation off

again. The Liquino is again in the basic condition.


Advertising