2 example of report - glp mode – Metrohm 711 Liquino User Manual

Page 131

Advertising
background image

4. Detailed description

711 Liquino, Instructions for use

125

4.9.2

Example of report - GLP mode

Validation of dosing units

Parameter Report

711 Liquino 0220 711.0011

date 1998-10-03 time 11:32 - 11:35

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: GLP user method: GLP test

operator D. Miller

drive id Dosino1

dos. unit id Dos20_1

parameters

>GLP parameter

dosing drive: DOS2

dos.unit: default

temperature 22.5 °C

>GLP options

liquid: water

weight from: balance

balance id

>GLP reports

report1: result

report2: parameter

report3: curve

report4: detail

report5: off

report6: off

report7: off

>>err. limits 20mL-cyl.

error max.vol.+/- 60 µL

slope +/- 0.003

intercept +/- 30 µL

============


← instrument number and program version

← date and time

← 3-line report header

← mode and method name

← request texts

← parameters


← options

← report definitions






← configured error limits


← identification for reproduced reports



Result Report

711 Liquino 0220 711.0011

date 1998-10-03 time 11:32 - 11:35

*** Dev. Dept.

*** dm

*** Metrohm Ltd

mode: GLP user method: GLP test

operator D. Miller

drive id Dosino1

dos. unit id Dos20_1

dosing unit size 20 mL

temperature 22.50 °C

liquid water

>evaluation

intercept (<30) -11.2 uL ok

slope (+/-0.003) 1.00076 ok

error max. vol. (<60)-2 uL ok

-> GLP test PASSED!

============


← instrument number and program version

← date and time

← 3-line report header

← mode and method name

← request texts

← parameters

← results (with limits)

← result

Advertising