3.2 Tutorial

711 Liquino, Instructions for use

30

3.2.6

PIP – precise pipetting/diluting


Example: Pipet 1mL and dilute with 9mL. Distilled water is used as the

"hydraulic“ liquid.


Preconditions / Preparation

The easiest and most accurate means of pipetting is with pipetting attach-
ment 6.1562.040. The keyboard can be used to proceed from step to step.

Define a dosing unit which corresponds to the tubing of the Dosino used.
This procedure is described on page 22. (This is not necessary if the pipet-
ting attachment is used as this corresponds to the standard settings.)

Method editing

*

MODE

-

X

DOS NR

SELECT

P

The PIP mode must first be activated. This sets all method parameters

to standard values. Press the <MODE> key and then <SELECT> until
‘PIP’ appears in the display:

Di spl ay:

mode sel ect : PI P

ENTER

YES

Press <ENTER>. The instrument is now in the PIP mode.

Di spl ay:

PI P * * * * * * * *

r eady 0. 000 mL

PARAM

B

The <PARAM> key is used to access the parameter menu, where all

the possible settings for a PIP method can be found.

Di spl ay:

par amet er s

>PI P par amet er s

The submenu ‘PIP parameters’ contains the main settings for a PIP

method. Press <ENTER> to open the submenu.

Di spl ay:

>PI P par amet er s

pi pet t i ng vol . 1. 000 mL

Enter the required pipetting volume (1 mL) and press <ENTER> to

accept the entry.

Di spl ay:

>PI P par amet er s

di l ut i ng vol . 9. 000 mL

Enter the 9 mL used as diluting volume in the example and press

<ENTER>

QUIT

NO

In the submenu ‘PIP parameters’ no further entries are necessary.

Press the <QUIT> key to exit the submenu.

Di spl ay:

par amet er s

>PI P par amet er s

J

<ENTER>


The next submenu is accessed with <>.

Di spl ay:

par amet er s

>PI P dosi ng uni t

The submenu ‘PIP dosing unit’ is opened with <ENTER>.

Di spl ay:

>PI P dosi ng uni t