0 45.2 co – LaMotte DO 6 Dissolved Oxygen Meter User Manual

Page 24

Advertising
background image

5.4.2 Salinity Setting Adjustment

NOTE: This mode is available

only from the mg/L (ppm)
Concentration Set Up menu.
1. Repeat steps 1 to 5 as in the

steps of Pressure Setting
Adjustment.

2. Press the HOLD/ENTER key

twice till the meter displays the
Salinity setting mode “b.SAL”.

3. Press the HOLD/ENTER key

to enter the Salinity setting
mode.

4. Use the

p and q keys to enter

the salinity of your solution in
parts per thousand (ppt).
Factory default is 0.0. Salinity
values up to 50 ppt can be
entered.

5. Press HOLD/ENTER key to

confirm the value. A
confirmation indicator “CO”
will be displayed for 1.5
seconds before returning to the
main group menu “DPr.7”.

6. Press CAL key to return to the

measurement mode.

23

.sal

HOLD

ENTER

HOLD

ENTER

0.0

45.2

co

“CO” will display
for 1.5 seconds

Display will return back to “DPr.7”
subgroup menu

Figure 5: Salinity Setting Adjustment Sequence