7.6.2 User Reset
This program clears all data
including calibration and
other customized setup
functions and reset it to
factory default.
You can skip Calibration
Reset and proceed straight to
User Reset
From “rSt.5” menu,
1. Press HOLD/ENTER key
three times until meter
displays the User Reset
menu “b.USR”
2. Press HOLD/ENTER key
to enter the menu.
3. Use the
p and q keys to
activate or deactivate the
User Reset selection.
a.
YES – activates User
Reset
b.
NO – deactivates
User Reset
4. Press HOLD/ENTER key
to confirm the selection
made.
5. If selection is ‘YES’, meter would immediately return to measurement
mode. Otherwise it will return to the main group menu. Press CAL to
return to the measurement mode.
35
HOLD
ENTER
Figure 26: User Reset Selection