10 ERROR MESSAGES

LCD Display

Indicates

Cause

Solution

Low Battery
indicator

Low battery
level

Need new batteries or
battery connection is
bad.

Clean battery contacts.
Replace batteries with
fresh ones, noting
polarity.

"Err 1"

in %

Saturation
Mode

% Saturation
Calibration
error

Calibration is
performed when
factory calibrated
absolute value is out
of calibration range –
10.1% to 49.9%

Check the value of the
calibration solution. If
zero calibration is done,
make sure the limit of
10% is not exceeded.
If message persist,
recondition your probe.

"Err 1" in mg/L
(ppm)
Concentration
Mode

Concentration
Calibration
error

Calibration is
performed when the
calibration solution
factory absolute
measurement is below
2.00 mg/L or ppm

Check the calibrating
solution to be above 2
mg/L or ppm.
Check that the correct
temperature and
salinity setting has been
set prior calibration.

“UR”/”OR”
with blinking
“A”
annunciator in
Temperature
Mode

ATC probe
error, Under
Range, Over
Range

ATC probe is
removed or broken
while the meter has
the ATC feature
activated.

Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.

“——“with
blinking “A”
annunciator in
% Saturation
and
Concentration
Mode

ATC probe
error

ATC probe is
removed or broken
while the meter has
the ATC feature
activated

Check the probe’s
temperature input
phono jack connection
to the meter. Ensure
probe is not broken or
punctured.

+

-