LaMotte DO 6 Dissolved Oxygen Meter User Manual
Page 48

% Saturation Calibration
Points
100% in saturated air or
air-saturated water.
0% in zero oxygen solution
% Saturation Calibration
Limits
Factory calibrated absolute value of
10.0% and below for 0% point &
50% to 200% for 100% point.
Concentration Calibration
Window
+/- 40% from the factory default
measurement value. Minimum
reading allowed is 2.00 mg/L (ppm).
Temperature Calibration
Window
+/- 5°C from factory default
measurement
Offset Adjustments (%
Saturation)
+/- 10.0 of reading in Saturation
mode
HOLD function
Yes
Auto-Off function
Selectable Auto Off function. (20
minutes after last key press)
Display
Customs Single 4 Digit LCD
Inputs
BNC for DO & 2.5 mm Phono for
temperature
Operating Range
0 to 50 °C
Power Requirements
4 AAA-sized batteries (included)
Battery Life
> 700 hours (Alkaline Batteries)
Dimensions
Meter: 14 cm (L) x 7 cm (W) x 3.5
cm (H)
Probe: 115 mm (L) x 12 mm (Dia)
with 3-ft cable.
Probe’s membrane housing: 16 mm
(Dia)
47