12 SPECIFICATIONS
% Saturation Mode
Range
Resolution
Relative accuracy
0.00 – 200.0 %
0.1 %
± 1.5% of Full Scale
mg/L (ppm)
Concentration Mode
Range
Resolution
Relative accuracy
0.00 – 20.00 mg/L or ppm
0.01 mg/L; 0.01 ppm
± 1.5% of Full Scale
Temperature
Range
Resolution
Relative accuracy
-5.0 – 105.0 °C
0.1 °C
± 0.5 °C
Salinity Correction
Range
Resolution
Method
0.0 – 50.0 ppt
0.1 ppt
Automatic correction after manual
input
Barometric Pressure
Correction (mm Hg)
Range
Resolution
Method
500 to 1499 mm Hg or 66.6 to 199.9
kPA
1 mm Hg or 0.1 kPA
Automatic correction after manual
input
Automatic Temperature
Compensation
0.0 to 50.0°C
Manual Temperature
Compensation
0.0 to 50.0°C
Probe (DO / Temp)
Galvanic / Thermistor
Probe Diameter
Body 12 mm, Cap 16 mm
Response Time
60 seconds to achieve 95% of the
reading