3.2

Movable Electrode Assemblies

Use the nylon cleaning brushes included with the PROTEAN IEF cell or a soft-bristled brush to
clean the movable electrode assemblies with hot water and Liqui-Nox. Take care to ensure the
platinum wire does not break. Rinse with DI water and air dry. Do not use ethanol to wash or dry
the movable electrode assemblies.

3.3

Cups

The Cup Loading Tray's cups are designed to be disposable and are not recommended for
reuse.

*Alconox and Liqui-Nox are registered trademarks of Alconox.

Section 4
General Guidelines for Sample Preparation

See ReadyStrip IPG Strip instruction manual for general guidelines for sample preparation. For
additional information on sample preparation, go to www.ProteomeWorksSystem.com