Problem Cause

Solution

3. Sample wicks

The sample cups are

For optimum performance use new

out of sample

disposable. When re-using

sample cups for each focusing run.

cups.

the sample cups, salts
can build up in the
corners and provide a
capillary path for the
sample to travel.

4. Programmed

Final combined conductivity

Run at maximum attainable voltage for

max. voltage is

of the IPG strip, sample,

longer time to arrive at the required

not reached

and rehydration buffer

number of volt-hours.

with a current

determines the maximum

limit of 50 µA

attainable voltage.

If possible, reduce the conductivity of

per strip.

the rehydration buffer by reducing the
Bio-Lyte

®

concentration.

Increase the current limit per strip.
However, higher current limits generate
more heat and increase the possibility of
burned regions in the IPG strips.

Section 6
Warranty

The Cup Loading Tray is warranted for 1 year against defects in materials and workmanship. If
any defects should occur during this warranty period, Bio-Rad Laboratories will replace the
defective parts without charge. However, the following defects are specifically excluded:

Defects caused by improper operation

Repairs or modifications performed by anyone other than Bio-Rad Laboratories or their
authorized agent

Damage caused by accidental misuse

Damage caused by disaster

Common replacement parts including platinum wire

Damage caused by use of organic solvents or basic detergents

For inquiries or to request repair service, contact your local Bio-Rad office.