Bio-Rad PROTEAN® IEF Cell Accessories User Manual

Page 6

Advertising
background image

Figure 3. Movable electrode assemblies.

1.4

Specifications

Material

Polycarbonate

IPG strip size capacity

7 cm–24 cm

Tray Capacity

Up to 12 IPG strips per run

Electrodes

Platinum wire on movable electrode assembly

Inter-electrode distance User defined. Dependent on IPG strip length
IPG strip rehydration

Rehydration/equilibration tray specific to IPG strip length

Section 2
Instructions

2.1

General Overview

Isoelectric focusing with the Cup Loading Tray includes the following steps:

1. Prepare rehydration solutions and rehydrate IPG strips overnight in rehydration tray with

appropriate volume of rehydration buffer.

Note: The Cup Loading Tray is not intended for passive or active IPG strip rehydration. Please
use appropriate rehydration/equilibration tray.

2. Prepare sample.

3. Transfer the rehydrated IPG strips from the rehydration tray and place the strips gel side up

into the focusing tray’s channels.

Note: Reduced rehydration times result in incomplete rehydration of the IPG Strips causing an
imperfect seal between the sample cup and the IPG strip.

4. Optional. Place electrode wicks at anode and cathode.

5. Position the movable electrode assemblies on top of the strips at the anode and cathode

ends. Ensure the spring clips are in contact with the electrode contact areas.

6. Place the sample loading cups on the rehydrated IPG strips near the anode for basic pH

range and near the cathode for acidic pH range IPG strips.

7. Load sample in sample cup and overlay with mineral oil.

8. Overlay IPG strips in each channel with mineral oil.

9. Transfer tray to cooling platform in PROTEAN IEF cell and start isoelectric focusing.