Bio-Rad CHEF Mapper® XA System User Manual

Page 24

Advertising
background image

Flg 4.2. Multi state angle example.

For example, to mimic a two state run, combine two vectors oriented at half the includ-

ed angle on either side of the vertical (combinations of states are called blocks). This is
equivalent to a two state run with a 96° included angle.

A combination of states designed as a unit is called a block. There can be a total of eight blocks
per run, with each of these blocks composed of 1–15 states. Section 9.7 gives examples of
multi state runs.

When MULTI STATE is pushed, the following screen is displayed:

Intrp =: This prompt is asking if the program is to have interrupts (secondary pulses). Enter
0 if no interrupts are to be used and 1 if interrupts will be used. Interrupts are explained in
Section 9.3.

Program with Interrupts? [ ]

180

°

-

+

0

°

20

+48°

-48°

96°

180°

180°

State 1

State 2

State 1 + State 2