After State 1 parameters are entered, the screen will ask if another State is needed.

Continue with another State?: To program another state, press 1, then ENTER. If another
state is not desired, press 0, then ENTER. To program a second state in the same block, enter
the appropriate values at the prompts.

Blk1 St02: You are now in Block 1, State 2. Continue with up to 15 states (vectors) for each
block. To program a different block, press 0 and ENTER when asked to Continue with
another State?
The screen will appear as:

Continue with another Block?: To enter another block, press 1, then ENTER. Up to eight
blocks can be programmed. If another block is not desired, press 0, then ENTER. The system
will display the message: A Program is in memory. Please enter another command. At this
point the program can be initiated by pressing START RUN.

For example, second block prompts would appear:

Blk2 Runtim: Enter the total run time for Block 2, then ENTER. The allowable run time
range is 1 minute–999 hours. Enter the run time as one time unit, either hours or minutes.
When entering the run time, put in the numeric value, then press HOURS or MINUTES.

Note: The prompt for interrupts (Intrp) in Blocks 2–8 is active only if interrupts were
chosen for Block 1 (1= yes at Intrp prompt in Block 1).

Blk2 RunTim = [

] Intrp = [

]

Continue with another Block? [

]

Blk2 St02 [

. ]

]

Ang = [

[

.

. ]

Continue with another State (Vector) ? [

]