Method 2: Routine Cleaning –– Protein Analysis

Use this procedure to clean the electrode manifold in the Experion electrophoresis
station immediately after each run
.

Warning: Never store the cleaning chip inside the electrophoresis station. Store the empty
cleaning chip covered to keep the wells clean.

1.

Fill a cleaning chip with 800 µl DEPC-treated water

10

. Gently tap the side of the cleaning

chip to remove any trapped bubbles from the wells.

2.

Open the lid of the electrophoresis station and place the cleaning chip on the chip
platform.

3.

Close the lid and leave it closed for 1 min.

4.

Open the lid and remove the cleaning chip.

5.

Allow the electrodes to dry for 1 min.

6.

Close the lid.

7.

Replace the water in the cleaning chip after use to avoid contamination. For storage,
remove the water from the cleaning chip and store the chip in a clean location.

10

DEPC-treated water is not required for this cleaning application; however, always use ultrapure (at minimum 0.2 µm-filtered)
water. Do not use autoclaved water.