Bio-Rad Bio-Plex Pro™ RBM Canine Kidney Toxicity Assays User Manual
Page 20
![background image](/manuals/600334/20/background.png)
18
3. Click Format Plate and format the plate according to the plate layout
created in Section 1 (Plan Plate Layout). To modify the plate layout,
follow the steps below (see Figure 4).
a. Select
the
Plate Formatting tab.
b. Select the standards icon
S
and drag the cursor over all
the wells that contain standards. Repeat this process for
blanks
B
, controls
C
, and samples
X
.
4. Click Enter Standards Info in the Protocol Settings bar.
a. Enter the highest concentration of each analyte in the top row
(labeled S1) of the table. S1 concentration information is listed in
the product data sheet.
b. Enter a dilution factor of 3 and click Calculate. The concentrations
for each standard point will be populated for all analytes in the table.
Fig. 4. Plate formatting.