Plan plate layout, Plan plate layout 8 – Bio-Rad Bio-Plex Pro™ RBM Canine Kidney Toxicity Assays User Manual

Page 10

Advertising
background image

8

1. Plan Plate Layout

Determine the total number of wells in the experiment using the Plate Layout
Template on page 28 or the Plate Formatting tab in Bio-Plex Manager

software. A suggested plate layout is shown in Figure 2, with all conditions
in duplicate.

1. Assign standards to columns 1 and 2, with the highest
concentration in row A and the lowest concentration in row H.

2. Assign the blank to wells A3 and A4. The blank should consist of

standard diluent alone and be processed in the same manner as
sample and standard wells. Note that Bio-Plex Manager automatically
subtracts the blank (B) MFI value from all other assay wells.

3. User-specific controls, as well as the quality controls supplied in the

kits, are assigned to wells in columns 3 and 4.

4. The remainder of the plate is available for samples.

Legend

S Standards

B Blank

X Samples

C Controls

Fig. 2. Suggested plate layout. For detailed instructions on
plate formatting in Bio-Plex Manager, see section Read Plate.

Advertising