Bio-Rad Bio-Plex Pro™ RBM Canine Kidney Toxicity Assays User Manual

Page 29

Advertising
background image

27

Possible Solutions

If samples contain little or no analyte,
negative values observed may
be due to statistical variation. If
assay drift is suspected, retest the
samples by positioning them next
to the standards. If contamination
of standards is suspected, check
the standard replicate value and
be careful when adding samples to
the wells. Matrix effects could also
produce negative sample values.
Bio-Plex Manager

software

automatically subtracts the blank (B)
FI value from all other assay wells.
While this has no impact on observed
concentrations of samples within the
assay working range, it may result
in a negative FI value if the blank’s
FI value is greater than either the
standard or the sample value. If this is
undesirable, then reformat the blank
wells as sample (X) or control (C) in
the protocol or results file.

Check if any interfering components
such as heparin-based anticoagulant,
additives, or gel from separators
were introduced into the samples.
Avoid using hemolyzed and
heavily lipemic samples. Remove
visible particulate in samples by
centrifugation. Avoid multiple freeze-
thaw cycles of samples.

Possible Causes

Impact of Sample Matrix

Negative MFI values in samples

Poor precision in serum and
plasma sample measurements

Advertising