11

Setting up a Vacuum Manifold

Calibrate the vacuum manifold by placing a standard 96-well flat bottom
plate on the unit and adjusting the pressure to –1 to –3” Hg. In general,
100 µl liquid should take 3–4 sec to clear the well. For more detailed
instructions, refer to bulletin #10005042.

4. Prepare Reagents

1. Reconstitute the following lyophilized reagents in dH

2

0 before use

according to the table below.

a. Allow vial to sit at room temperature for a minimum of 5 min, not to

exceed 30 min.

b. Mix by vortexing at a medium setting.

2. Bring the 10x assay buffer to room temperature (RT).
a. Mix by inversion to ensure all salts are into solution.
b. Prepare 1x assay buffer — dilute 1 part 10x assay buffer with

9 parts of dH

2

0.

Reagent

dH

2

O Volume

Standards mix

150 µl

Control 1

100 µl

Control 2

100 µl

Blocking buffer

1.5 ml

Standard diluent

1.0 ml

Detection antibodies

4.8 ml

Table 4. Reagent volume.