Removing Outliers
Outliers are identified as standard data points that do not meet accuracy
or precision requirements and should be considered invalid when
performing curve fitting. As such, they should be removed to generate a
more realistic and accurate standard curve. This may result in an extended
assay working range and allow quantitation of samples that might
otherwise be considered out of range (OOR).

In Bio-Plex Manager software version 6.0 and higher, outliers can be
automatically removed by selecting the Optimize button in the Standard
Curve window. In Bio-Plex Manager software 6.0 and earlier versions,
outliers can also be manually selected in the Report Table. Visit online Help
to learn more about the standard curve optimizer feature and how outliers
are determined.

Previous Versions of Bio-Plex Manager Software

For instructions on using previous versions of Bio-Plex Manager software,
please contact Bio-Rad Technical Support.