Section 10

Troubleshooting Guides

This troubleshooting guide addresses problems that may be encountered with

Bio-Plex

®

Precision Pro™ cytokine assays. If you experience any of the problems

listed below, review the possible causes and solutions provided. This will assist
you in resolving problems directly related to how the assay steps should be
performed. Poor assay performance may also be due to the Bio-Plex array
reader. To eliminate this possibility, we highly recommend use of the Bio-Plex
validation kit. This kit will validate all the key functions of the array reader and
assist the user in determining whether or not the array reader is functioning
properly.

Possible Causes

Possible Solutions

20

High Inter-Assay CV

Standards were not
reconstituted consistently

Incubate the reconstituted
standards for 30 min on ice. Always
be consistent with the incubation
time and temperature.

Reconstituted standards and
diluted samples were not stored
properly

Reconstituted standards and diluted
samples should be prepared on ice
as instructed. Prior to plating, the
reconstituted standards and diluted
samples should be equilibrated to
room temperature.

High Intra-Assay CV

Bottom of filter plate not dry

Dry the bottom of the filter plate with
absorbent paper towel (preferably
lint-free) to prevent cross-
contamination.