Bio-Rad Bio-Plex® Precision Pro™ Human Cytokine Assays User Manual

Page 25

Advertising
background image

23

Possible Causes

Possible Solutions

High Background Signal

Incorrect buffer was used
(for example, assay buffer
used to dilute standards)

Use sample matrix or serum
standard diluent to dilute
cytokine standards.

Spiked “0 pg/ml” wells by mistake

Be careful when spiking standards.
Do not add any antigens in the 0
(blank) point.

Streptavidin-PE incubated
too long

Follow the procedure incubation
time.

Poor Recovery

Expired Bio-Plex reagents were
used

Check that reagents have not
expired. Use new or unexpired
components.

Incorrect amounts of components
were added

Check your calculations and be
careful to add the correct volumes.

Microplate shaker set to an
incorrect speed

Check the microplate shaker speed
and use the recommended setting.
Setting the speed too high may
cause splashing and contamination.
Use the recommended plate shaker.

Pipetting technique

Pipet carefully and slowly when
adding standards, samples,
detection antibodies, and
streptavidin-PE, especially when
using a multichannel pipet. Use a
calibrated pipet. Change pipet tip
after every volume transfer.

Advertising