Bio-Rad Bio-Plex® Precision Pro™ Human Cytokine Assays User Manual
Page 24
![background image](/manuals/600338/24/background.png)
22
Possible Causes
Possible Solutions
Low Bead Count
Miscalculation of bead dilution
Check your calculations and be
careful to add the correct volumes.
Beads clumped in multiplex
bead stock tube
Vortex for 15–20 sec at medium
speed before aliquoting beads.
Vacuum on for too long when
aspirating buffer from wells
Do not apply vacuum to the filter
plate for longer than 10 sec after the
buffer is completely drained from
each well.
Did not shake filter plate enough
before incubation steps and prior
to reading
Shake the filter plate at 1,100 rpm
for 30 sec before incubation steps
and immediately before reading
the plate.
Reader is clogged
Refer to the troubleshooting guide
in the Bio-Plex hardware
instruction manual.
Low Signal or Poor Sensitivity
Standards reconstituted incorrectly Follow the cytokine standard
instructions carefully.
Detection antibody or
streptavidin-PE diluted incorrectly
Check your calculations and be
careful to add the correct volumes.