Bio-Rad BioLogic Sample Loading Loops User Manual

Page 15

Advertising
background image

d.

In the Inject Sample area of the window, select the Injection Buffers, the buffer
composition, the flow rate of the BioLogic DuoFlow workstation pumps, and the
sample volume to be injected onto the column.

The BioLogic DuoFlow system will now automatically control the loading and injection
of the sample. Note that the rinse function is not available when the DynaLoop is being
used.

4. Continue writing your desired separation protocol.

The AVR7-3 Injection Valve Function

During the run, the AVR7-3 injection valve functions as follows (Figure 6):

Load. While in this position the valve connects ports 5 and 4 for equilibration of the
column and for sample elution. In this position, sample loop is loaded to the desired
volume via port 2 and buffer is expelled from the dynamic loop through port 1.

Inject. While in this position the valve connects ports 5 and 6 and ports 3 and 4 for
applying the sample onto the column. The flow from the workstation pump forces the
sliding piston to expel the sample onto the column.

Purge. While in this position the valve connects ports 5 and 7 and allows purging or
buffer changes of the workstation pump without the need to remove the column from
the system.

Section 4
Care and Maintenance

4.1 Troubleshooting

Table 1. Troubleshooting the DynaLoop

Problem

Possible Cause

Corrective Action

High backpressure

Plugged filter

Replace filter

when loading or injecting

(Section 4.3)

Leakage at the compression

Loose or worn ferrules

Tighten or replace ferrules

screw end fittings

Leakage into the plastic

Worn O-rings on

Replace O-rings

shield

sample or buffer end

(Section 4.4)

11

Advertising