Bio-Rad Model EG-1 Econo™ Gradient Monitor User Manual

Page 15

Advertising
background image

12

Larger or smaller than

Chart recorder input

Check chart recorder

expected gradient

voltage set incorrectly

settings (See Sections

profile

4.2 and 5)

If using Gradient

Re-calibrate the

mode, buffer A and

gradient mode

B values not set
correctly

Ghost peaks appear in

Charged sample

No problem

gradient profile

detected (e.g. protein)

Air bubbles passing

Use degassed buffers.

through flow cell

Check for loose plumb-
ing connections

Chart recorder

Chart recorder is not

Check connections (See

not responding

connected properly

Sections 3.2 and 4.2)

Full scale of recorder

Set recorder range to

set too high

1 V

Recorder has bottomed Readjust pen position
out

Flow cell not

Check flow cell

connected in-line

plumbing

No power

Faulty power

Check power cable

connection

connection

Blown fuse

Check fuses (See
Appendix A)

Unit not switched on

Check power on/off
switch on rear panel

Conductivity

Dirty flow cell

Clean flow cell (See

measurements with

Section 6)

same buffer appear
to decrease over time

Problem

Possible Cause

Solution

M7318150H 2/27/1 2:06 PM Page 12

Advertising