Bio-Rad Model EG-1 Econo™ Gradient Monitor User Manual

Page 17

Advertising
background image

Appendix C
Technical Specifications

General

Operating mode

Conductivity, in units of Siemens (See
Section 1)

Operating temperature

4-40 °C

Sensitivity ranges (full scale) 0-10µS, 0-100µS, 0-1000µS, 0-10mS
Output signal

0-1 Volt

Power requirements

90-264 V, 50-60 Hz (depending upon
power adaptor used)

Power consumption

20W maximum

Noise

Short Term: < 0.1 µS
Long Term: < 0.5 µS

Linearity

+/-0.5 % full scale

Accuracy

+/-2 % full scale

Drift

< 0.5 µS/hour/°C

Dimensions

Control Unit:

15.60 x 22.86 x 8.89 cm (W x L x H)

Flow Cell:

3.1 cm diameter by 3.8 cm length

Weight

Control Unit: 600g

Case Material

Chemical and fire resistant polypropylene

Regulatory

CSA: C22.2 No. 151-M1986

Compliance:

TÜV: EN61010-1:1993
CE: EN55011: 1991 Class A

EN50082-2:1995

Flow cell

Type

Flow through, externally mounted

Temperature compensation

0-60 °C

Internal volume

8 µl, swept volume

Cell constant

Approximately 40*

Maximum operating pressure 60 psi, 4 bar, or 0.41 mPa
Cable length

1.22 M

Wetted Parts

Titanium, Kel-F and Tefzel

Fittings

Removable luer (Will also accept 1/4”-28
flat bottom fittings)

* Most applications do not require an exact cell constant. If it is necessary

to determine the exact cell constant, Bio-Rad‘s Conductivity Standard,
catalog number 167-0445, can be used.