Section 16
Troubleshooting Guide

Sample is too viscous

High concentration of host
nucleic acids in lysate

Insufficient amount of
homogenization buffer

Viscosity of extract can be
reduced by addition of
Benzonase nuclease
(1.7 units/ml) with 1 mM
MgCl

2

to fragment

bacterial DNA. Incubate on
ice for 15 min

Dilute sample by adding
more homogenization
buffer

Sample application

causes column to clog

Insufficient clarification of
sample

Reduce cell debris from
clogging the column by
increasing the
centrifugation speed or
filtering the sample

No protein is eluted

Expression of target
protein in extract is very
low and is not found in the
eluate

Check expression level of
protein by estimating the
amount in the extract,
flow-through, eluted
fraction, and pellet upon
centrifugation. Use western
blotting with anti-6xHis
antibodies, target-protein-
specific antibodies, ELISA,
or enzyme activity
determination

Apply larger sample
volume

Minimize contact with
hydrophobic surfaces (that
is, polystyrene tubes).
Proteins at low concen-
tration may bind to the
surface of the tube

Problem

Possible Cause

Solution