No protein is eluted

(continued from

previous page)

Elution conditions are too
mild or protein may be in
an aggregated or multimer
form

Elute with pH or an
imidazole step-elution

Protein precipitates

during purification

Temperature is too low

Aggregate formation

Perform the purification at
room temperature

Add solubilization agents
to the samples and/or
buffers: 0.1% Triton X-100,
Tween-20, 20 mM
β-mercaptoethanol and
≤20% glycerol to maintain
protein solubility

Poor recovery of target

protein

Protein is found in the
flow-through

Binding capacity of the
column has been
exceeded

Target protein was not
detected in the flow-
through

Strong nonspecific
adsorption of the target
protein to the matrix

See recommendations in
“No protein eluted” section

Increase the column size
or reduce the sample
volume application

Capillary sample loop is
too small

Reduce hydrophobic
adsorption by including
detergents, or organic
solvents, or by increasing
the concentration of NaCl

Problem

Possible Cause

Solution