CANON Digital Galvano Scanner System

KP-1SM Series + GC-251

Users Manual 1.00

69

Parameter ID

10

Parameter

Name

Encoder Periodicity

Data

Encoder periodicity (Unit: pulse)

Return Value

0: Parameter setting successful
1: Parameter setting unsuccessful

Explanation

This parameter is used to set the number of pulses per rotation of the motor encoder.
KP-1SM30,KP-1SM100: 83328 pulses
Whenever the motor model is changed, the setting of this parameter should be
changed.

Related
Command
/Parameter

Parameter ID

11

Parameter

Name

Sampling Time

Data

Sampling period (Unit: ns)

Return Value

0: Parameter setting successful
1: Parameter setting unsuccessful

Explanation

This parameter is used to set the sampling interval of the controller.
The usual value is 10000 for 10 μs.

Note: This parameter usually requires no change.

Related
Command
/Parameter

Parameter ID

12

Parameter

Name

Origin Clearance Timing (Head 1)

Data

Origin clearance timing (0 or 1)

Return Value

0: Parameter setting successful
1: Parameter setting unsuccessful

Explanation

This parameter is related to origin detection.
The value checked by “Command ID = 101: Counter Clear Timing” is set.
Each motor has a unique value.

Note: An appropriate value is set at shipping. When only the motor is replaced, the
value should be checked and set by a command.
The Origin Clearance Timing (Head 2) parameter should be set at the same time.

Related
Command
/Parameter

Counter Clear Timing (Command ID = 101)
A1C101/ 0 (Axis 1 Head 1)
Set the value checked above.