CANON Digital Galvano Scanner System

KP-1SM Series + GC-251

Users Manual 1.00

74

Parameter ID

26

Parameter

Name

Raster Scan Time

Data

Scan Time (Unit: See Explanation.)

Return Value

0: Parameter setting successful
1: Parameter setting unsuccessful

Explanation

This command is used to set the raster scan time (operation by command).
The scan time is the set value × 10 μs.
E.g. For the interval of 100 ms, set 10000 (10000 × 10 μs = 100 ms).


Related
Command
/Parameter

Raster Scan Duty Ratio (Parameter ID = 27)
Raster Scan Angle (Parameter ID = 28)
Raster Scan Interval Time (Parameter ID = 30)
Raster Scan Start Position (Parameter ID = 31)

Parameter ID

27

Parameter

Name

Raster Scan Duty Ratio

Data

Raster scan duty ratio (Unit: %)

Return Value

0: Parameter setting successful
1: Parameter setting unsuccessful

Explanation

This parameter is used to set the duty ratio of raster scan (operation by command).
(See Explanation of Parameter ID = 26.)

Related
Command
/Parameter

Raster Scan Time (Parameter ID = 26)
Raster Scan Angle (Parameter ID = 28)
Raster Scan Interval Time (Parameter ID = 30)
Raster Scan Start Position (Parameter ID = 31)




CW

CCW

Scan Angle

0

Raster Scan Time

Scan Angle

CW

CCW

Scan Angle

0

Raster Scan Time

Scan Angle