2-7

Functional Description and Specifications

Rear Panel (Input/Output)

The 2000-810-01 Indicator rear panel is shown in Figure 2-5.

Figure 2-5: 2000-810-01 Rear Panel

A — SOLENOID CONTROL (Hirose RM15TRD-12S)

Not used (optional)

B — ANALOG OUTPUT (BNC)

Analog output provides a voltage level output that is directly
proportional to transducer input, from -1.8V (full scale counter
clockwise) to +1.8V (full scale clockwise). Zero offset +/-100
mV. Linearity +/-1% of reading. Maximum load = 5 Ma.

When using the 4-in-1 transducer, zero output
fluctuates until one of the transducers is scan selected
by applying 5% of full scale torque.

C — PRINTER SELECTOR

The INTERNAL/EXTERNAL switch disables the built-in printer.

A

B C

D

E

F

G