Dial , beam or screw driver mode – CDI Torque MULTITEST Motorized Loader User Manual

Page 88

Advertising
background image

6-16

2000-800-02 Motorized Loader

DIAL , BEAM OR SCREW DRIVER MODE:

( This mode is used to check or test the Dial wrenches ).

· Setup the Torque Tester:

** Mode:- Peak or TRACK
** Select an appropriate Unit (ft-lb, Nm, etc.)

· Setup the Motorized Control Box:

** Press

F1

key to select Dial Mode.

The LCD display reads:

Press

RUN

key to start Dial Mode:

The LCD Display reads:

Use the cursor keys Up, Down, Left, Right to enter
Target Value, Number of Cycles and Direction.

Typically, the Target Value in Dial Mode should be
about -5% to -10% of the required torque.

Example :

For Dial wrench check at 55 ft-lb, set Target Value at 50 ft-lb.

Press RUN or Enter key to accept all Parameters.

NOTE :

The default Setup upon Power up is :
* Target :

7% of Transducer.

* Cycle:

00001

* Dir:

CW

The MCB does NOT accept any Target value that exceeds
Max Transducer Range.

*** MODE (Dial) ***

Press Run to Start

*** DIAL MODE ***

Target Cycle Dir
50.00 ft-lb 00005 CW

Advertising