Archived 6/1/10 – ETS-Lindgren 95252-1 Bulk Current Injection Probes (Archived) User Manual
Page 12
Advertising
![background image](/manuals/636325/12/background.png)
Conducted Emission
BULK CURRENT INJECTION PROBES
8
© ETS-Lindgren, L.P. – February 2005
Rev C – PN 399263
Calibration Jig 95241-1 may be used for this determination.
Typical transfer impedance values for the Model 95242-1 Probe
when used as a sensor are:
Frequency
(MHz)
Z
T
(
Ω)
2
7
4
14
6
21
8
26
10
30
15
35
20
37
30
38
40
39
50
39.5
60
40
80
41
100
41
140
42
180
42
220
43
260
43
300
43
320
43.5
340
43.6
360
44
380
42.5
400
39
420
32
440
25
460
17
Table 2
High sensitivity and minimal core gap indicate that the current
probe should be used only on signal lines where heavy currents are
not encountered. Heavy currents may affect measurement
accuracy should current probe core saturation occur.
Archived 6/1/10