Conducted Emission

BULK CURRENT INJECTION PROBES

8

© ETS-Lindgren, L.P. – February 2005

Rev C – PN 399263

Calibration Jig 95241-1 may be used for this determination.

Typical transfer impedance values for the Model 95242-1 Probe

when used as a sensor are:

Frequency

(MHz)

Z

T

(

Ω)

2

7

4

14

6

21

8

26

10

30

15

35

20

37

30

38

40

39

50

39.5

60

40

80

41

100

41

140

42

180

42

220

43

260

43

300

43

320

43.5

340

43.6

360

44

380

42.5

400

39

420

32

440

25

460

17

Table 2

High sensitivity and minimal core gap indicate that the current

probe should be used only on signal lines where heavy currents are

not encountered. Heavy currents may affect measurement

accuracy should current probe core saturation occur.