Archived 6/1/10 – ETS-Lindgren 95252-1 Bulk Current Injection Probes (Archived) User Manual

Page 18

Advertising
background image

Specifications

BULK CURRENT INJECTION PROBES

14

© ETS-Lindgren, L.P. – February 2005

Rev C – PN 399263

SERIES SPECIFIC ELECTRICAL SPECIFICATIONS

Electrical

Specifications

95236-1

95242-1

95252-1

Frequency

Range

10 kHz to 100 MHz

2 to 400 MHz

10 MHz to 1000 MHz

Maximum Input

Power

100 W

200 W (See Note)

100 W

Maximum Input

Current

20 Amperes

60 Amperes

25 Amperes

Maximum Core

Temperature

80 degrees C

80 degrees C

80 degrees C

Recommended

Maximum

Temperature

Rise

35 degrees C

35 degrees C

35 degrees C

Maximum Time

for Continuous

Rating at Full

Power

30 minutes

30 minutes

15 minutes

Turns Ratio

1:2

1:1

1:1

Inductance

47

µ

H, ±20%

0.8

µ

H, ±20%

0.5

µ

H, ±20%

Table 6

Note:

The power limit of the 95236-1 probe is 100 W. Powers in excess of 75 W should be used with

care to avoid excessive temperature in the equipment under test. Prolonged testing should be

avoided, particularly if unattended.

Note:

The power limit of the 95242-1 probe is 200 W. Powers in excess of 100 W should be used

with care to avoid excessive temperature in the equipment under test. Prolonged testing should be

avoided, particularly if unattended.

Note:

The power limit of the 95252-1 probe is 100 W. Powers in excess of 75 W should be used with

care to avoid excessive temperature in the equipment under test. Prolonged testing should be

avoided, particularly if unattended.

Bulk Injection

Probe

Useful Range

Insertion Loss (Typical)

dB

Range

95236-1

10 kHz—100 MHz

<-15 dB

-35 db ± 3 dB

1 MHz-100 MHz

@ 10 kHz

95242-1

2 MHz—400 MHz

<-15 dB

10 MHz – 400 MHz

95252-1

400 MHz—1 GHz

<-12 dB

<-22 dB

<-30 dB

400 MHz—700 MHz

750 MHz—900 MHz

950 MHz—1 GHz

Table 7