Application, Conducted susceptibility, Archived 6/1/10 – ETS-Lindgren 95252-1 Bulk Current Injection Probes (Archived) User Manual

Page 6

Advertising
background image

Application

BULK CURRENT INJECTION PROBES

2

© ETS-Lindgren, L.P. – February 2005

Rev C – PN 399263

APPLICATION

The principal use of the bulk current injection probe is for inducing

relatively large RF currents into the signal and power circuits of

equipment under test for conducted susceptibility. A secondary

application would be to use the same probe in a more familiar role

as a sensor for measuring weak conducted RF currents.

CONDUCTED SUSCEPTIBILITY

Conducted susceptibility testing is intended to insure that RF

signals, when coupled on to interconnecting cables and power

supply lines of an IUT, will not cause malfunction or degradation

of performance. In addition, this testing can provide an amplitude

vs. frequency malfunction signature for the system which, when

compared with the levels of current on the cables in a typical

operating environment, can assist in the determination of adequate

safety margins.

TYPICAL TEST SETUP

Typical conducted susceptibility tests require that all power and

interconnecting cables be tested by subjecting them to the required

current or voltage levels, while monitoring the applied current

using a current probe. Usually, a reference level calibration is

performed using a calibration jig with a specified impedance. This

reference curve is then replayed to expose the IUT to a controlled

stress level, while a current probe is used to insure that a low

impedance IUT is not overstressed.

Note: Some tests may allow the reference calibration to be

performed at a lower level and then scaled up to the required

power level when applied to the IUT.

Archived 6/1/10

Advertising