Precautionary Measures

BULK CURRENT INJECTION PROBES

12

© ETS-Lindgren, L.P. – February 2005

Rev C – PN 399263

PRECAUTIONARY MEASURES

CAUTION: RF fields can be hazardous. Observe appropriate RF

exposure limits.

When measuring conductors that are not insulated, use extreme

care when installing the current probe and taking measurements. If

possible, de-energize the test sample during assembly and

disassembly of the setup. Also, arrange to center the test

conductor in the current probe window for additional voltage

breakdown protection.

Do not permit the un-insulated current probe connector and cable

connectors to come in contact with the ground plane or other

nearby conductors. This will prevent possible measurement error

due to ground loops, and will avoid danger from high voltages.

Ensure that the 50 Ohm load is capable of safely dissipating the

incurred power. Should the load become disconnected, the

developed voltage will be come much greater and may be very

dangerous.