Ear temperatures – Welch Allyn Spot Vital Signs LXi - User Manual User Manual
Page 49

Directions for Use
Temperature Operation
39
Ear Temperatures
Spot LXi with the Braun ThermoScan PRO 4000 thermometer takes a temperature in the
ear. The thermometer probe shape prevents insertion far into the ear canal which could
perforate the tympanic membrane.
To take an ear temperature reading:
1.
Pull the bottom of the thermometer gently toward you to remove it from the housing.
2.
Locate the probe cover box inside the thermometer housing. Firmly push the probe
tip into the probe cover box. When the probe cover is in place the thermometer turns
on automatically.
3.
Listen for the ready signal beep and three dashes to appear on the thermometer
display.
4.
Fit the probe snugly into the ear canal and then push the Start button.
a.
If the probe is positioned correctly in the ear canal the «ExacTemp» light flashes.
When the thermometer detects an accurate measurement, the «ExacTemp» light
is continuously on, a long beep signals the end of the measurement, and the
display shows the result.
b.
If the probe is positioned incorrectly in the ear canal or is moved during the
measuring process, the «ExacTemp» light goes out, a sequence of short beeps
sounds, and the display shows an error message («POS» = position error).
5.
Press the ejector button to eject the used probe cover into the trash when you are
finished taking the temperature.
6.
Return the thermometer to its holder. Spot LXi displays the patient’s temperature and
temperature scale in the Temperature Display area on the LCD (see “Display
Window” on page 22).
WARNING Keep the probe window clean, dry, and undamaged at all times to
ensure accurate measurements. To protect the probe window, always keep the
thermometer in the storage cover while transporting or when not in use.
WARNING Only use Braun ThermoScan probe covers with this thermometer.
Using other manufacturer’s probe covers or no probe cover may produce
temperature measurement errors and/or inaccuracies. If the thermometer is used
without a probe cover attached, clean the lens (see “Braun ThermoScan PRO
4000 Thermometer” on page 60).
WARNING Do not autoclave.
WARNING The thermometer is not waterproof. Do not immerse or drip fluids on
it. Should this occur, dry the thermometer with warm air. Check for proper
operation and accuracy.