Troubleshooting, Error codes, 8 - troubleshooting – Welch Allyn Spot Vital Signs LXi - User Manual User Manual
Page 55

8
45
Troubleshooting
Error Codes
The following tables of error codes provide a quick reference of the descriptions and
probable causes of error codes. For service-level troubleshooting, refer to the Service
Manual (part number 704432).
Table 12. General Error Codes
Code
Description
Corrective Action
C12
Device outside operating temperature
range.
Change ambient temperature.
C13
Low battery level.
Charge battery.
E30
Internal malfunction.
Contact Technical Service.
E31
Internal malfunction.
Contact Technical Service.
E32
Internal malfunction.
Contact Technical Service.
E33
Internal malfunction.
Contact Technical Service.
E38
Date and time not set.
Set date and time (see “Date/Time Menu Options” on
page 30)
E42
Internal malfunction.
Contact Technical Service.
E44
Internal malfunction.
Contact Technical Service.
E45
Internal malfunction.
Contact Technical Service.
Table 13. Blood Pressure Error Codes
Code
Description
Corrective Action
C01
Blood pressure reading cancelled by user.
Retake blood pressure reading.
C02
Unable to release cuff pressure.
Check tubing for kinks and connection integrity.
C03
Inflation too quick.
Check tubing and connections.
C04
Air leak.
Check blood pressure cuff and tubing connections.
C05
Unable to determine blood pressure.
Check connections; restrict patient movement.
C06
Unable to determine blood pressure.
Check connections; restrict patient movement.
C07
Internal NIBP error.
Device will power down.
E10
Cuff pressure limits exceeded.
Device will power down.
E11
Cuff pressure duration exceeded.
Device will power down.
E20
Internal NIBP error.
Device will power down.