Troubleshooting, Error codes, 8 - troubleshooting – Welch Allyn Spot Vital Signs LXi - User Manual User Manual

Page 55

Advertising
background image

8

45

Troubleshooting

Error Codes

The following tables of error codes provide a quick reference of the descriptions and
probable causes of error codes. For service-level troubleshooting, refer to the Service
Manual (part number 704432).

Table 12. General Error Codes

Code

Description

Corrective Action

C12

Device outside operating temperature
range.

Change ambient temperature.

C13

Low battery level.

Charge battery.

E30

Internal malfunction.

Contact Technical Service.

E31

Internal malfunction.

Contact Technical Service.

E32

Internal malfunction.

Contact Technical Service.

E33

Internal malfunction.

Contact Technical Service.

E38

Date and time not set.

Set date and time (see “Date/Time Menu Options” on
page 30)

E42

Internal malfunction.

Contact Technical Service.

E44

Internal malfunction.

Contact Technical Service.

E45

Internal malfunction.

Contact Technical Service.

Table 13. Blood Pressure Error Codes

Code

Description

Corrective Action

C01

Blood pressure reading cancelled by user.

Retake blood pressure reading.

C02

Unable to release cuff pressure.

Check tubing for kinks and connection integrity.

C03

Inflation too quick.

Check tubing and connections.

C04

Air leak.

Check blood pressure cuff and tubing connections.

C05

Unable to determine blood pressure.

Check connections; restrict patient movement.

C06

Unable to determine blood pressure.

Check connections; restrict patient movement.

C07

Internal NIBP error.

Device will power down.

E10

Cuff pressure limits exceeded.

Device will power down.

E11

Cuff pressure duration exceeded.

Device will power down.

E20

Internal NIBP error.

Device will power down.