Specifications, Performance, Blood pressure accuracy – Welch Allyn Spot Vital Signs LXi - User Manual User Manual
Page 61: 9 - specifications

9
51
Specifications
Performance
This section describes normal ranges for Spot Vital Signs LXi.
Blood Pressure Accuracy
Blood pressure accuracy meets or exceeds AAMI SP10:2002 standards for non-invasive
blood pressure accuracy (± 5 mmHg mean error, 8 mmHg standard deviation). Blood
pressure accuracy is validated for pressure measurement using the upper arm only.
Cuff Pressure Range
0 to 300 mmHg
Systolic Range
60 to 250 mmHg
Diastolic Range
30 to 160 mmHg
Blood Pressure Determination Time
Typical: 15 seconds
Mean Arterial Pressure Range
40 to 190 mmHg
Pulse Rate Range
(using SpO
2
determination)
25 to 240 bpm
Pulse Rate Range
(using Blood Pressure determination)
35 to 199 bpm
Pulse Rate Accuracy (using SpO
2
determination)
Without Motion: 25 to 240 bpm ± 3 digits
1
With Motion: normal physiologic range (55 to 125 bpm) ± 5 digits
Low Perfusion: 25 to 240 bpm ± 3 digits
1
Pulse Rate Accuracy
(using Blood Pressure determination)
±5.0%
Overpressure Cutoff
315 mmHg ±15 mmHg
1
Specification applies to device performance and was validated with Biotek and Nellcor simulators.