Guidance and Manufacturer’s Declaration

Emissions and Immunity Information

Electromagnetic Emissions

The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs
LXi should assure that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment - Guidance

RF emissions

CISPR 11

Group 1

The Spot Vital Signs LXi uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic
equipment.

RF emissions

CISPR 11

Class B

The Spot Vital Signs LXi is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/
flicker emissions

IEC 61000-3-3

Complies

Electromagnetic Immunity

The Spot Vital Signs LXi is intended for use in the electromagnetic environment specified below. The customer or user of the Spot Vital Signs
LXi should assure that it is used in such an environment.

Immunity Test

IEC 60601
Test Level

Compliance Level

Electromagnetic Environment - Guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.

Electrical fast transient/
burst

IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output
lines

±2 kV for power supply
lines

±1 kV for input/output
lines

Mains power quality should be that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be that of a typical commercial or
hospital environment.