Chapter 11 Specifications
191
Table 35. Setup window
—
Arrhythmia Alarms
Features
Customizable default
Factory default
Alarm level
Medium for Pause, off for all
other arrhythmia types except
VFib, asystole and V-Tach
Lethal for VFib, asystole, V-Tach
Arrhythmia alarm limits
Presets values
Presets values
V-Tach, 125; PVC run, 6; PVC/
min, 30; Tachycardia, 180;
Bradycardia, 45; Irregular Limit,
45
Print on alarm
Off for all
Options:
Irregular Limit
Single ECG (active with 5-wire cable
only)
Pacer Display
Analyze Pacers
Off
Off
Off
Off
PaceMaker NonCapture
Disabled
Leads used for arrhythmia analysis
Welch Allyn 1500 Patient Monitor,
Micropaq, and Propaq LT monitor 5-wire
cables:
•
Arrhythmia channel 0 uses lead II.
•
Arrhythmia channel 1 uses lead V.
•
Arrhythmia channel 2 uses lead III.
Propaq Encore, Propaq CS:
The leads that the user selects for ECG1
and ECG2 are used for arrhythmia
analysis.
Arrhythmia alarms for paced patients
Arrhythmia alarms depend on Acuity
System pacemaker detection vs. monitor
pacemaker detection.