ST Analysis option
The Acuity System ST Analysis option analyzes the ST segment of current ECG beats
relative to the ST segment of a reference beat. Specifically, the system reports mm or mV
of deviation between the current beat’s ST-segment offset and the reference beat’s
ST-segment offset.
The system learns the reference beats when a learning or relearn function occurs. For a
listing of when learning or relearn functions occur, see
In the ST Analysis Setup window, you can do the following things:
•
Specify analysis preferences.
•
Set alarm limits.
•
Set ST alarming priority level.
•
Set Print on Alarm.
•
View current beats and reference beats.
•
View current deviation values, including alarm values.
•
Initiate relearn of a patient’s reference beats for ST analysis only.
ST analysis is not performed in these circumstances:
•
If Analyze Pacers is set on in the patient’s Arrhythmia Alarms Setup window.
•
If the patient monitor is set in neonate mode.
•
During the occurrence of arrhythmia events.