68

Chapter 4 Adjust monitoring settings and patient information

Welch Allyn Acuity Central Monitoring System

WARNING Carefully monitor lead-wire connections of patients prone to
arrhythmia events. If a lead wire disconnects during an event, the Acuity
System’s ability to accurately detect the event can be compromised.

WARNING Minimize patient movement during a RELEARN, as movement
creates noise during the reading. If the system learns an abnormal rhythm or
learns a rhythm during noise, it designates the learned rhythm as normal.
Thereafter, rhythms of this type might not trigger an alarm. Ensure that the
patient remains still for 5 minutes after the RELEARN before allowing them to
move.