NIBP algorithm

To modify the NIBP algorithm used on the device:

1. From the Advanced screen, use Up or Down navigation buttons to highlight NIBP

algorithm.

2. Press the Select button to edit the entry.
3. Use the Up or Down navigation buttons to highlight SureBP or Step.
4. Press the Select button to accept the entry as shown.

If SureBP is selected, the device inflates the cuff to the appropriate level, measuring the blood

pressure as the cuff is inflating. The systolic display shows the pressure in the cuff as the

blood pressure determination is in process. If Step is selected, the device will inflate the cuff,

then measure the blood pressure while deflating the cuff. If the device is unable to determine

a blood pressure while the cuff is inflating due to patient movement, excessive noise, or an

arrhythmia, the device will use the Step algorithm to inflate the cuff to a higher pressure, then

attempt to measure the blood pressure while deflating the cuff.

The default NIBP algorithm is SureBP.

Language

To change device language:

1. From the Advanced screen, use Up or Down navigation buttons to highlight Language.
2. Press the Select button. The Language screen displays.
3. Use the Up or Down navigation buttons to highlight the correct Language option.
4. Press the Select button to choose a language.

A popup confirmation screen will display your language selection.

The default language is English.