No cuff inflation – Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual

Page 70

Advertising
background image

Possible cause

Corrective action and explanation

recommends deflation no faster than 3 mmHg

per second.

Only use a sphygmomanometer that is calibrated.

An uncalibrated sphygmomanometer may result

in inaccurate blood pressure measurements.

Poor auscultatory sound recognition by observer

Use high-quality stethoscope. Have a different

observer check patient’s blood pressure.

Cuff inflation and deflation with no blood pressure reading

displayed

Possible cause

Corrective action and explanation

Leak in pneumatic system

Inflation too quick; check NIBP cuff and

tubing connections may display.
Cuff pressure limits exceeded. Powering

down may display.
NIBP air leak; check cuff and tubing

connections may display.
Unable to determine NIBP; check inflation

settings may display.
Ensure all cuff attachments are tight.
Carefully check for leaks in the blood pressure cuff,

tubing, and pressure hose attached to ProBP 3400.

Arm movement during cycle

Unable to determine NIBP; check

connections; limit patient movement may

display.
Keep arm still during blood pressure cycle.
Movement may cause inaccuracies from artifact, long

cycle times, and error message.

Cuff tubing or pressure hose movement artifact

Unable to determine NIBP; check

connections and tubing for kinks may display.
Do not contact cuff tubing or pressure hose during

blood pressure cycle.
Movement may cause inaccuracies from artifact.

User may have pressed the

Blood Pressure

Start/Stop

button

User cancelled NIBP reading will display. Press

the

Blood Pressure Start/Stop

button to

initiate blood pressure measurement.

No cuff inflation

Possible cause

Corrective action and explanation

Connections between device and cuff loose

Check all connections.

64 Troubleshooting

Welch Allyn Connex® ProBP™ 3400 digital blood pressure device