Specifications, Physical specifications, Mechanical specifications – Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual
Page 57

Specifications
Physical specifications
Performance
This section describe normal ranges for the ProBP 3400 device.
Blood Pressure Accuracy
Blood pressure accuracy meets or exceeds ANSI.AAMI SP10:2002 standards for noninvasive blood
pressure accuracy (±5 mmHg mean error, 8 mmHg standard deviation). Blood pressure accuracy is
validated for pressure measurement using the upper arm only.
Cuff pressure range
0 to 300 mmHg
Systolic range
60 to 250 mmHg
Diastolic range
30 to 160 mmHg
Mean Arterial Pressure (MAP) range
MAP is a calculated reading that yields an approximate
value.
40 to 190 mmHg
Pulse rate range
35 to 199 bpm
Pulse rate accuracy
±5.0%
Overpressure cutoff
300 mmHg ±15 mmHg
Mechanical specifications
Dimensions
Height: 15.0 cm (5.91 inches)
Width: 8.0 cm (3.15 inches)
Depth: 5.6 cm (2.20 inches)
Weight
450 g (0.99 lb)
Mounting
Custom mobile stand
Custom wall mount
Custom desk mount
Portability
May be used as a handheld device
51